WATER RESOURCES RESEARCH, VOL. 45, W00D23, doi:10.1029/2008WR007052, 2009

for

Full Article

Environmental temperature sensing using Raman spectra DTS fiber-optic methods Scott W. Tyler,1 John S. Selker,2 Mark B. Hausner,3 Christine E. Hatch,1 Thomas Torgersen,4 Carl E. Thodal,5 and S. Geoffrey Schladow6 Received 31 March 2008; revised 29 May 2008; accepted 16 October 2008; published 28 January 2009.

[1] Raman spectra distributed temperature sensing (DTS) by fiber-optic cables has

recently shown considerable promise for the measuring and monitoring of surface and near-surface hydrologic processes such as groundwater–surface water interaction, borehole circulation, snow hydrology, soil moisture studies, and land surface energy exchanges. DTS systems uniquely provide the opportunity to monitor water, air, and media temperatures in a variety of systems at much higher spatial and temporal frequencies than any previous measurement method. As these instruments were originally designed for fire and pipeline monitoring, their extension to the typical conditions encountered by hydrologists requires a working knowledge of the theory of operation, limitations, and system accuracies, as well as the practical aspects of designing either short- or long-term experiments in remote or challenging terrain. This work focuses on providing the hydrologic user with sufficient knowledge and specifications to allow sound decisions on the application and deployment of DTS systems. Citation: Tyler, S. W., J. S. Selker, M. B. Hausner, C. E. Hatch, T. Torgersen, C. E. Thodal, and S. G. Schladow (2009), Environmental temperature sensing using Raman spectra DTS fiber-optic methods, Water Resour. Res., 45, W00D23, doi:10.1029/2008WR007052.

1. Introduction [2] The use of heat as a tracer in hydrologic systems has a long and successful history. Both the temporal and spatial distribution of fluid temperature (either as a surrogate for or input into energy conservation formulations) has provided process insights from deep in the crust to the atmosphere. Tracking thermal pulses, estimating fluid fluxes, tracing surface water – groundwater exchange, and predicting groundwater recharge have all benefited from the study of thermal signatures (the reader is referred to Anderson [2005] and Stonestrom and Constantz [2003] for thorough reviews and analyses of thermal methods in near-surface hydrology). Measurement systems have ranged from handheld mercury thermometers to forward looking infrared imagery and typically provide either a ‘‘point-in-time’’ measurement across a landscape or borehole or, alternatively, a ‘‘pointin-space’’ measured over a time span. As a result, thermal measurements are often spatially or temporally constrained, and few approaches are available to provide coverage both in space and in time. 1 Department of Geological Sciences and Engineering, University of Nevada, Reno, Nevada, USA. 2 Department of Biological and Ecological Engineering, Oregon State University, Corvallis, Oregon, USA. 3 Graduate Program of Hydrologic Sciences, University of Nevada, Reno, Nevada, USA. 4 Department of Marine Sciences, University of Connecticut, Groton, Connecticut, USA. 5 U.S. Geological Survey, Carson City, Nevada, USA. 6 Department of Civil and Environmental Engineering, University of California, Davis, California, USA.

Copyright 2009 by the American Geophysical Union. 0043-1397/09/2008WR007052$09.00

[3] The Raman spectra distributed temperature sensing (DTS) for use in hydrologic applications presents an opportunity for continuous interrogation of streams, catchments, lakes, the atmosphere, and the oceans. These temperature-sensing systems were first developed in the 1980s [e.g., see Dakin et al., 1985; Kurashima et al., 1990] and were utilized primarily for fire monitoring, pipeline monitoring, and other industrial applications. While reports of DTS applications in geothermal wells first appeared in the mid-1990s to late 1990s [Hurtig et al., 1994; Sakaguchi and Matsushima, 2000], it is only since 2006 that hydrologic applications have employed dedicated installations rather than opportunistic use of existing communication cables [Selker et al., 2006a, 2006b; Westhoff et al., 2007; Lowry et al., 2007; Henderson et al., 2008; Moffett et al., 2008]. For example, Figure 1a shows a time series of vertical temperature measurements taken through the thermocline at Lake Tahoe in northern California on 7 June 2007. At each time slice (1 min), over 400 temperature measurements were made through the water column by suspending a fiber-optic cable and interrogating it from the surface. Figure 1b shows only the upper 100 m of the profile to highlight the thermocline interface. The oscillations that can be seen in the thermocline (for example, between 30 and 40 m depth) at a frequency of 10 3 Hz may be the result of internal gravity waves traversing the lake after several days of high winds. The internal gravity waves propagate along the thermocline interface, where the vertical gradient in water density (due to surface heating) is greatest. With a 1-min integration time, the temperatures are resolved to 0.1°C over each meter of fiber. Unlike thermistors or thermocouple strings suspended at fixed depths sampling in time or conductivity-temperature-depth casts at high vertical

W00D23

1 of 11

W00D23

TYLER ET AL.: DTS FIBER-OPTIC SENSING

W00D23

resolution but without repeating in time, DTS offers an unparalleled spatial and temporal-sampling resolution for many hydrologic applications. [4] Previous work has thoroughly described the theoretical basis and physical design of commercially available DTS systems, and the reader is referred to Selker et al. [2006a] for more details. DTS is not, however, without complications and requires significant initial investment and careful experimental design to produce at its capacity. The primary objectives of this work are (1) to provide the readership with knowledge of the practical aspects of DTS and DTS experiments in hydrologic applications and (2) to assess limitations and uncertainties that must be addressed to fully utilize these systems for hydrologic measurement. The manuscript begins with a review of DTS theory, followed by an assessment of the accuracy and precision of the instruments that can be expected for hydrologic applications, and concludes with practical experimental design considerations for the deployment of the instrument-cable system. Application of DTS in environmental sciences is a rapidly evolving field, and we hope that this discussion will provide guidance for the use of DTS in hydrologic applications.

2. Assessing DTS Systems for Hydrologic Measurement

Figure 1. Time series of water column temperatures collected from a 420-m fiber-optic cable suspended from a research buoy to the bottom of Lake Tahoe (430 m) on 7 June 2007. (a) The entire depth profile displaying a strong thermocline at 20– 40 m and constant (in time and depth) profile of bottom waters. (b) An expanded view of the upper 100 m of the profile, suggesting a record of temperature fluctuations in the thermocline (clearly visible at 25– 35 m depth) consistent with internal gravity waves generated from a series of storms several days before the measurements were made. (c) The time evolution of temperature at a depth of 27 m below the water surface, displaying a periodicity of 25 min.

2.1. DTS Instrument Considerations [5] Effective application of DTS to environmental sensing requires appreciation of the physics of the measurement, which provide context and criteria for instrument and cable selection, deployment design, and calibration methods. Because these choices determine the quality of the data collected by the DTS instrument during the experiment, they must be made in light of the goals of the experiment, the theoretical and physical limitations of the technology, and the physical limitations of the field deployment, including cost. [6] To measure temperature in the fiber, the instrument pulses a laser and times the return signal. Finite measurement speed and dispersion of light along the fiber imply that Raman spectra DTS systems measure the average temperature along a length of the fiber-optic cable, typically 1 –3 m, rather than at a single point. Commercial DTS systems use a variety of signal generating and processing schemes, including time domain, frequency domain, and spread spectrum modulation. With the speed of light in the fiber known, it is possible to calculate the source of the Raman scatter within 1 m for cables up to 5 km and on the order of 1.5 m for cables up to 10 km in length because of the intrinsic dispersion of the multimode fibers employed in the method. Achieving this spatial resolution requires truncation of the signal. For example, most commercial fibers have an index of refraction close to 1.5, so the speed of light in the fiber is about 0.2 m/ns. Since the backscattered arrival includes the two-way travel time, the delay from the instant of light injection to observation of the backscattered arrival is approximately 10 ns/m, with the backscatter from any particular meter arriving spread out over a 5-ns period. To avoid temperature distortions due to dispersion of light within the fiber, the first and last 1 – 2 ns of the pulse return must be trimmed. This leads to greater signal loss for finer

2 of 11

W00D23

TYLER ET AL.: DTS FIBER-OPTIC SENSING

spatial resolution: assuming a 1-ns trimming, 1-m resolution loses 40% of the signal (2 ns of the 5-ns signal), while 2-m resolution would lose only 20% (2 ns of a 10-ns signal). The bandwidth capacity of the fiber itself can also affect the spatial-resolving ability of DTS. In addition, the sampling rate of instruments (limited by the circuit-processing speeds used by each instrument) can lead to additional signal loss at high spatial resolution. A longer spatial-averaging interval can significantly improve the quality of the DTS temperature measurements, so spatial averaging should always be selected with care in accordance with the experimental requirements. [7] DTS temperature measurements employ the amplitude ratio of the amplitudes of the backscattered Stokes to the anti-Stokes signals [Selker et al., 2006a]. The Stokes and anti-Stokes signals are the result of the Raman effect, in which incident light interacts inelastically with the electrons in the molecular bond. The molecule will emit light (scatter) light at two characteristic frequencies (Stokes referring to light of lower frequency and anti-Stokes returning light of higher frequency), reflecting the quantum energy states of the electrons which the photons encountered. The higher the temperature, the greater will be the number of electrons in high-energy states, increasing the fraction of anti-Stokes scattering relative to the Stokes signal. The ratio of these signals provides a quantity independent of light intensity that depends only on the temperature of the fiber at the location identified by the two-way travel time of light since the time of injection. Since only a small fraction of the incident light is scattered in an optical fiber, signal strengths are very low, which is the primary limit on the precision of DTS measurement. Greater signal strength may be achieved by longer integration time (summing the ratios obtained from a large number of pulses), brighter lasers, or spreadspectrum modulation. Temperature resolution approaching 0.01°C is possible with long integration times if the DTS device itself is protected from changes in temperature [Selker et al., 2006a]. [8] The time-averaging interval is also an important consideration in experimental design. Fundamentally, the precision of the method is a function of the Stokes to antiStokes backscatter ratio precision. Since this factor is proportional to the number of photons collected by the detectors (linear in time), the precision of this ratio follows the central limit theorem and thus is proportional to the square root of time. All other things being equal, a longer time-averaging period will provide a more precise temperature measurement. [9] Hydrologists need to understand the anticipated repeatability and accuracies that can be expected from DTS in order to adequately design, logistically plan, and financially budget installations. As DTS systems measure temperature as a function of both space and time, it is important to clearly define repeatability, accuracy and resolution. Taylor and Kuyatt [1994, p. 14] define repeatability as ‘‘closeness of the agreement between the results of successive measurements of the same measure and carried out under the same conditions of measurement.’’ To evaluate repeatability, Taylor and Kuyatt [1994] suggest the same measurement procedure, observer, instrument, and location. Furthermore, the measurements must be repeated over a short period of time. By statistically examining the

W00D23

dispersion of the repeated measurements, the repeatability of the measurements can then be quantified. [10] Repeatability is important for hydrologic applications; however, absolute accuracy, or the ability to measure a temperature accurately, is also an important factor in designing a DTS deployment. Absolute temperature accuracy is typically referenced to a standard measurement method, such as standard platinum resistance thermometers. For most commercial DTS systems, manual calibration against such standard thermometers can easily be used to develop accuracies of ±0.1°C or better. Typical DTS calibration procedures are discussed further in this work. 2.2. Repeatability [11] We have chosen to adapt the concept of Taylor and Kuyatt [1994] for our typical conditions and divide repeatability into two components: temporal and spatial. Temporal repeatability may be an important limiting factor in the accuracy of long-term DTS system deployments. Many hydrologic experiments rely upon long time series data records, in which the evaluation of a time-varying signal leads to conclusions about a given hydrologic system. The temporal repeatability (including instrument drift) can be evaluated by monitoring a section of fiber held at constant temperature (for example, in an ice bath at 0°C or any known and unvarying temperature environment) and subjecting the instrument to environmental stresses such as varying instrument temperature, as would typically be found in field installations. Spatial repeatability quantifies the ability of a system to correctly interpret a sequence of temperatures in space. For example, when DTS systems are deployed in streams to detect groundwater inputs [Selker et al., 2006b; Westhoff et al., 2007; Lowry et al., 2007], it is necessary to understand the instrument’s ability to detect the spatial patterns of stream temperature and its reliability to do so. [12] A series of controlled field trials in the deep waters of Lake Tahoe were conducted in June 2007 to assess both the temporal and spatial repeatability of several DTS systems. Three manufacturers supplied DTS instruments for the testing (Agilent Technologies, Inc.; Sensornet, Ltd.; and SensorTran, Inc.). Data presented here are not designed to evaluate the merits of one manufacturer over another, as the optimal operating conditions are dependent not only on the DTS instrument but also on the software used and on the fiber-optic cable. As such, the instrument numbering (1 – 3) does not necessarily correspond to the order given above. These data are instead shown to illustrate the range of repeatability and spatial resolution that is commercially available today. For these tests, the three DTS instruments were brought aboard the University of California, Davis research vessel John LeConte and were deployed at a monitoring buoy located in the north portion of Lake Tahoe in 430 m of water. A loop of BruSteel1 fiber-optic cable (Brugg Kable AG, Brugg, Switzerland) containing two multimode fibers encased in a sealed stainless steel capillary and covered by wire rope and a PVC jacket was lowered to within 10 m of the bottom and was looped back to the surface. This fiber, along with all fibers used in the testing program reported here, was terminated with E2000 connectors, the general industry standard for fiber-optic connections for DTS applications.

3 of 11

W00D23

TYLER ET AL.: DTS FIBER-OPTIC SENSING

W00D23

Figure 2. Field repeatability test results of three DTS systems in the deep waters of Lake Tahoe, California: (a) time series from a single depth (400 m) and (b) average (30 min) vertical distribution of temperatures over a deep portion of Lake Tahoe, where temperatures were unchanging. [13] The temperature profiles taken during this experiment show a thermocline approximately thirty meters below the surface (Figure 1a), with near-constant temperature below the thermocline. Thermistor measurements (Figure 2) taken at a depth of approximately 302 m show a constant water temperature of 5.002° ± 0.002°C over a 30-min period just prior to the deployment of the fiber-optic cable and DTS measurements. Taking advantage of this constant temperature of the deep lake water, it is possible to demonstrate both the temporal and spatial repeatability. 2.2.1. Temporal Repeatability [14] Figure 2a shows the temperature time series for the three different instruments taken at a depth of 400 m. The time series was collected within a 2-h period, during which there was no change in the independent thermistor-measured temperature at these depths. Because of the small diameter of the cable, it quickly reached thermal equilibrium with the surrounding fluid. Instruments 1, 2, and 3 showed mean temperature readings of 5.33°, 5.18°, and 5.14°C, respectively, slightly higher than the independent thermistor measurement. This was likely due to the calibration bath used for each of the DTS instruments, which only had 5 m of cable immersed in an ice bath. The short calibration section was the result of logistical challenges on the vessel and would generally produce a calibration offset that was neither

accurate nor appropriate for each instrument. An ideal calibration uses a longer length of cable, typically 30 m or more, at a known constant temperature to determine the offset of the temperature measurements. [15] Despite the inaccurate calibration, the temporal repeatability of the three systems is quite good, with temporal standard deviations of 0.08°, 0.13°, and 0.31°C, respectively, for the three instruments shown. The temporal repeatability was consistent throughout the deep waters for each of the instruments tested on this particular fiber-optic cable but was somewhat elevated when the instruments were connected to a ‘‘well-used’’ cable that contained several step attenuations resulting in less signal returning to the sensors. The impacts of cable flaws/damage on data quality will be discussed in subsequent sections. [16] It is important to note that these tests were conducted over fairly short time durations (2 h) and that the long-term stability of the measurements was not tested. We have noted larger (±1° to 2°C) instrument drift when DTS instruments are deployed for multiday experiments and, in particular, when the instruments are subjected to significant internal temperature variations. The long-term stability can be improved using a dynamic calibration method available on some DTS instruments, and the importance and source of long-term drift is the subject of ongoing work.

4 of 11

W00D23

TYLER ET AL.: DTS FIBER-OPTIC SENSING

Figure 3. Illustration of the spatial-resolution calculation for a single trace (with instrument 2 on 4 June 2007 at 1640:07 local time) from the laboratory testing described. 2.2.2. Spatial Repeatability [17] The spatial repeatability of measurements was examined along a length of fiber at uniform temperature from traces taken over a specified time interval (we define a trace to be a single temperature versus distance graph). Spatial repeatability is an important consideration in streamgroundwater studies, where changes downstream in water temperature are used to infer groundwater inputs. It is much easier, for example, to immerse a long section of fiber in a well-mixed constant temperature bath than to maintain that bath temperature constant over long periods of time, particularly under field conditions. Spatial repeatability is quantified by taking the standard deviation of a number of consecutive measurements over a given length of fiber at a constant temperature during a single trace of the DTS instrument. [18] Figure 2b shows the 30-min averaged temperatures of the bottom 100 m of the lake for each of the tested instruments. Once again, the absolute accuracy was affected by the less than ideal temperature bath; however, the spatial repeatability for all three instruments is quite good with spatial standard deviations of 0.02°, 0.04°, and 0.08°C, respectively. The high-frequency spatial variability visible (sawtooth pattern) in the signal of instrument 3 may be partially the result of this instrument’s sampling resolution/ spatial resolution algorithm. It is clear, however, from Figure 2b that small changes in temperature over space can be easily resolved with DTS. 2.3. Resolution 2.3.1. Spatial Resolution [19] The spatial resolution of a DTS system is also an important consideration for hydrologists and refers to the spatial integration scale over which a single value of temperature is reported. Most DTS instrument manufac-

W00D23

turers state a minimum spatial resolution (on the order of +1 m), and this resolution can be adjusted upward in many instruments if coarser-scale measurements are more appropriate. In testing the spatial resolution of a DTS system, the length of fiber required to display a known step change in temperature along the fiber is determined. A simple laboratory setup was constructed using a single ice bath to examine the defined spatial resolution twice for each instrument (into the bath and out of the bath representing two ‘‘steps’’). We define in this work the ‘‘spatial resolution’’ of the DTS system as the length of fiber required to show 90% of the known temperature change (note that other definitions of spatial resolution commonly reported use a definition of 80% of the temperature change). The temperature change is calculated from the difference of the two reference temperatures. For example, if the bath temperature change is from 25° to 0°C, the spatial resolution is the distance over which the DTS temperature measurements change from 23.75° to 1.25°C, i.e., 90% of the total temperature change centered on the mean of the two constant temperatures. Figure 3 shows one example trace of the spatial resolution testing in which the fiber is exiting an ice bath. [20] The spatial resolution test will be affected by the spatial-averaging design of each DTS instrument, and the spatial resolution test is intended to quantify these effects. To ensure that the test is accurate, a length of fiber must be maintained at a constant temperature on each side of the step change. We recommend a minimum length of 10 times the spatial-averaging interval of the DTS system being tested. [21] A total of 60 time slices for each instrument was used to evaluate the spatial resolution. The fiber-optic cable used in the experiment used connectors that showed higher losses than that normally expected (higher than normal connection loss results in less return signal and lower resolution). Using the defined 90% step change, instrument 1 demonstrated a mean spatial resolution (over 60 tests) of 3.35 m with a standard deviation of 0.50 m; instrument 2 showed a spatial resolution of 1.91 ± 0.27 m; and instrument 3 returned a spatial resolution of 3.45 ± 0.39 m. These spatial resolutions are 2– 4 times that reported by the instrument manufacturers (instrument 1, 1 m; instrument 2, 1.5 m; instrument 3, 1 m) and reflect both differences in laboratory conditions and testing methodology and the use of low-quality fiber and connectors in this particular test. These results suggest that spatial resolution of DTS systems needs to be tested for individual applications and specifically for those fiber-instrument combinations used in the field to adequately quantify the spatial resolution that can be obtained. 2.3.2. Temporal Resolution [22] The temporal resolution of a DTS system is typically controlled by the instrument’s ability to statistically measure the returning Raman energies, which are both low in amplitude. Most commercially available DTS systems can resolve fiber temperatures over 10-s intervals and are sufficient for many rapidly changing hydrologic systems. As the sampling time is increased, the resolved temperatures become more accurate. However, longer temporal averaging reduces the probability of detecting a rapidly changing temperature signal. Because postcollection averaging of

5 of 11

W00D23

TYLER ET AL.: DTS FIBER-OPTIC SENSING

short sampling intervals provides the same statistical reduction of variance as one longer sampling interval of the same time, one should always choose the longest averaging time that still assures sufficient temporal resolution to satisfy the environmental-monitoring objectives of the study. Given the very low cost of data storage, it is strongly advised to carry out postcollection averaging rather than averaging during data collection, since any averaging done by the instrument cannot be reversed. [23] Temporal resolution is also profoundly affected by the choice of fiber-optic cable. Small-diameter cables using low heat capacity coatings and strength elements will respond more quickly to environmental temperature changes than will heavily armored and large-diameter cables. For many of the fiber-optic cables deployed to date in hydrologic applications, the time scale for the cable to respond to thermal changes is still quite short (typically on the order of seconds to 1 min), and it is important to conduct simple step change heating/cooling experiments to insure that the fiber used is compatible with the anticipated time scales of environmental temperature change. 2.4. Cable Options and Instrument Operation, Design, and Performance [24] Researchers contemplating DTS systems for hydrologic experiments need to consider the installation as a total system which includes fiber-optic cable design, configuration, and environmental constraints. In this section, we focus on ‘‘downstream’’ aspects of DTS, i.e., those components of the system such as the fiber-optic cable and connectors that transmit signals and affect signal quality. 2.4.1. Cable Options [25] The package of properly protected optical fiber(s) is referred to as the cable, which will include tensile strength members, physical armoring protecting the enclosed fiber from impingement, and a water-tight barrier to eliminate all chemical contact with the glass. It is very common for a cable assembly to contain several fibers, both for redundancy and for larger bandwidth. There are many cables currently available on the market today, so it is important to consider the experimental environment and potential stresses on the cable before making an investment. Typically, less than 10% of the cost of a cable suitable for field deployment for environmental sensing is attributable to the fiber itself. The fiber must be protected from potential damage, often including debris (e.g., rocks in rivers or falling branches in forests), animals, and pressure (in deep water). The specific construction of cable containing fibers is important depending on the conditions in which the cable will be deployed. [26] Optical fibers used for Raman DTS typically consist of a 50-mm diameter glass core surrounded by a 37-mm glass cladding that creates a reflective boundary by way of a change in index of refraction. The transition in optical properties between the core and cladding glass can be a gradual graded index or abrupt step index. For sensing at temperatures below 150°C, graded index fibers are preferred because of lower optical dispersion, while at elevated temperatures, or for applications with many tight bends, the step index fiber should be used. [27] Immediately surrounding the fibers are protective materials (buffering) which fall into two categories: loose tube and tight buffered. The loose tube construction sur-

W00D23

rounds the fiber with a hydrophobic gel. The gel protects the fiber from exposure to water, as contact with water can lead to hydrogen contamination of the fiber, compromising the index of refraction step function in the fiber package and thus leading to light loss. In addition, the gel-filled capillary helps physically isolate the fiber from damage due to mechanical impingements. The loose tube fiber and gel are housed within capillaries made from plastic or stainless steel, capable of providing protection sufficient for pressures found on the ocean floor and vehicular passage over cables on hard surfaces. Tight-buffered fibers have an injection-molded plastic coating protecting the fiber. These coatings provide excellent control of hydrogen diffusion and protection from mechanical abrasion, but any force applied to the outside of the buffering results in strain on the glass fiber. In general, tight-buffered cables should be avoided for environmental temperature sensing applications, though acceptable data may be obtained with care. [28] Tensile stress must also be avoided, and two measures are typically taken. In loose tube buffering, the fiber is ‘‘overstuffed’’ so that if the cable experiences limited elongation ($10/m) and quite heavy (>30 kg/km) and more difficult to handle than a less robust cable; they will also have a greater thermal inertia than lighter cables and will thus respond more slowly to temperature changes in the environment. [29] As light passes down the fiber, is scattered, and returns to be detected, some light is lost or attenuated uniformly along the fiber. Attenuation is defined as the loss of signal per unit length of fiber and is a function not only of fiber construction but of wavelength of signal. Standard telecommunications fibers used for environmental DTS have quoted losses of 0.3 dB/km (at standard telecommunications frequencies) which can be quantified during DTS calibration to determine the appropriate correction factors. However, other factors can lead to much higher, localized attenuation, including tensile stresses, hydrogen gas penetration, air gaps in junctions, and bend angles exceeding the internal angle of refraction. Typically, fiber bend radii of less than 2.5 cm should be avoided, but for short cable lengths, this can be reduced. [30] Simple attenuation typically results in a noisier temperature signal, as less light is returning to the detector per unit time. However, as DTS temperature is calculated from the ratio of Stokes and anti-Stokes returning signal, attenuation that affects these two signals differently will lead to incorrect temperature calculations. Unfortunately, most attenuation processes are wavelength-dependent, and therefore, the difference in attenuation of these two signals, termed differential attenuation or DA, is a very important factor in the accuracy of DTS-derived temperatures. Fortunately, experimental geometry, internal DTS algorithms,

6 of 11

TYLER ET AL.: DTS FIBER-OPTIC SENSING

W00D23

Table 1. Comparison of Single- Versus Double-Ended DTS Measurements Location

Single-Ended sa (°C)

Double-Ended sb (°C)

Near ice bath (3.5 – 26.5 m) Far ice bath (1061 – 1084 m) Cable midpoint (531.5 – 556.5 m)

0.26 ± 0.05 1.2 ±0.18 0.39 ± 0.06

0.47 ± 0.08 0.65 ± 0.10 0.31 ± 0.05

a

Values are the mean and standard deviations of 87 single-ended traces. Values are the mean and standard deviations of 87 double-ended traces.

b

and even multiple laser sources can be designed to minimize and/or correct for differential attenuation. The ubiquitous presence of DA requires use of known temperature baths at each end of each cable section free of connectors or fusion splices. 2.4.2. Instrument Options and Performance [31] An important decision for DTS cable deployment related to the impacts of differential attenuation is whether to use single- or doubled-ended measurements. Each has complementary advantages which may advocate the use of each for different experiments. Single-ended measurements refer to temperatures calculated from light transmission in only one direction in the fiber (analogous to a one-way, dead-end street). Single-ended measurements are most precise near the instrument and degrade with distance. Doubleended measurements are made by convolving single-ended measurements made from each end of a loop of cable with both ends of the fiber connected to the instrument. In a double-ended measurement, the temperature reported for each point incorporates data from traces taken in each direction, yielding greatest noise at the points closest to the instrument and least noise at the midpoint of the fiber. The advantage of the double-ended measurement is that it allows computation of the DA continuously along the cable for each measurement. For long-term monitoring or where the fiber has been or may be subjected to stresses/damage that lead to step or localized losses of light transmission, double-ended measurements will provide much higher quality data. Double-ended calibration typically requires only a temperature offset, whereas single-ended systems must be calibrated for differential loss, gain, and offset. Using a double-ended configuration, however, adds complexity to the system by requiring at least two input channels, twice the sensor length capacity and a doubling of the minimum measurement rate. [32] Double-ended measurement geometries should not be chosen automatically for all applications, however, as they tend to add noise to resolved temperatures nearest the DTS sensor. To demonstrate the impacts of these two measurement approaches, both single-ended and doubleended traces were collected from a 1000-m cable deployed along the shallow littoral zone of Lake Tahoe. At both ends of the installation, 0°C ice baths containing 20– 40 m of immersed cable were used first for calibration and subsequently for measurement of repeatability and accuracy. These reference baths were located close to the DTS instrument (termed ‘‘near ice bath’’) and at the distal end of the installed fiber (‘‘far ice bath’’). Table 1 shows the average standard deviation of temperature through both ice baths. The mean standard deviation for the far bath in single-ended mode was almost 5 times greater than the near

W00D23

bath, while in double-ended mode, the near and far baths’ standard deviation differed only slightly. However, it is important to note that the standard deviation of the near bath in the double-ended case is almost twice as large as that from the same bath in single-ended mode. The use of a double-ended measurement in this case has degraded the signal near the DTS unit. This is the result of combining the two-way signals, which distributes the noise more equally along the length of the fiber. [33] For applications where a uniform level of precision is needed along the length of the fiber, double-ended measurements will be preferred. If there is any chance of timedependent differential attenuation, the double-ended measurement can provide data that might otherwise contain artifacts that would not be possible to remove [see Tyler et al., 2008]. DTS instruments capable of double-ended measurement can typically be configured to alternate between single-ended and double-ended measurements, allowing for hybrid data collection. However, for sites where higher levels of precision are needed close to the instrument (for example, the data shown in Figure 1) or where it is not convenient to have a continuous fiber return to the DTS unit, single-ended measurements may be advantageous. An alternative approach to correct time-dependent DA is to add additional laser sources to the DTS instrument which specifically measure the loss as a function of frequency at the Stokes and anti-Stokes frequencies. Finally, several instrument manufacturers are now producing DTS systems that utilize two or more lasers tuned such that the Stokes frequency of one laser is very close in frequency to the antiStokes frequency of the second laser, and much of the differential attenuation can be eliminated. In all cases, it is therefore important for the researcher to carefully evaluate the environment, the most crucial locations of measurements, and the time scales of variation in order to most effectively deploy a DTS system. 2.4.3. Other Performance Factors [34] Numerous other factors can affect the accuracy and repeatability of the signal from a fiber-optic cable. Connectors and connection quality can have a significant impact on signal quality. The cable connection to the DTS instrument and to other cables may be achieved by a variety of methods that are dependent upon the installation. The fiberoptic industry has many standard connectors, with the E2000 connection having been adopted by the DTS industry. To minimize backscatter of the injected light, the connectors used for DTS must be angle polished, which is not standard for the multimode fiber required and thus must be special ordered. Mechanical connectors are susceptible to scratches, dust on the connecting faces, poor alignment, and air gaps which together can cause signal loss that may be a function of temperature and direction of light transmission. The asymmetrical losses are particularly problematic in that they are not well characterized in the interpretation of double-ended measurements and can give rise to temperature anomalies due to asymmetric DA. Manufacturers report that standard connectors typically cause a signal loss of 0.1 – 0.2 dB in the fiber line, although those losses are measured using a different wavelength of light than is used for DTS systems. The quality of a fiber-optic connector can be quite variable, depending on the precision of the fabrication process used to build it, as well as the age and

7 of 11

W00D23

TYLER ET AL.: DTS FIBER-OPTIC SENSING

W00D23

Figure 4. (a) Observed anomalies in reported temperatures at each connector (numbered points). (b) Magnitude of the returned signal (red, dimensionless) and of the signal attenuation (blue), which produces anomalous temperatures at each connector as a function of the number of connectors through which the light has passed. After eight connection passes, the signal amplitude of the returned signal is reduced such an extent that downstream temperatures cannot be resolved. cleanliness of the connector itself. Unfortunately, some E2000 connectors purchased from different manufacturers have even been found to be physically incompatible. [35] To assess the role of connector loss on data quality, a simple laboratory experiment was conducted through sequential connectors. A 21.5°C constant temperature bath was prepared, and a 30-m-long cable consisting of 12 fibers was submerged in the bath. At each end of the cable, connectors were used to join sequential fiber ends, resulting in a single cable effectively 360 m long with 12 connectors along its length. Results of this experiment are plotted in Figure 4. Each time the signal passes through a connector, there is a step loss in signal strength illustrated by the mean Stokes radiation (Figure 4b) averaged over the 10-min duration of the experiment. With each step loss in signal the variability of the calculated temperature in the constant temperature bath increases, with almost complete degradation of the calculated temperature after eight connectors (210 m in Figure 4a). As seen in Figure 4b, half of the original signal is lost after passing through only three connectors, and the noise introduced through multiple connectors overwhelms the temperature signal after seven or eight connectors. [36] Several alternative methods of mating fiber-optic cable for environmental applications are available. Mechanical splices use an index of refraction matching optical gel

to improve efficiency of light passage. Unlike the standard connectors, which require ends with connectors, mechanical splices can join bare, cleaved fibers. The closed splice clamps onto the buffer on each end of the fiber, providing strain relief for the exposed length of fiber inside it, and manufacturer’s specifications report expected loss of 0.05– 0.2 dB. However, we have found this low loss to be difficult to achieve in practice in the field, and such mechanical junctions have been seen to have temperature dependence in their DA, potentially leading to temperature offsets that are very difficult or impossible to fully eliminate. [37] Ideally, one would prefer to join fibers with a continuous column of glass, which is what a fusion splicer produces. Like mechanical splices, fusion splices require cleaved, stripped fiber, and the two cores must be aligned to within